Data Foundry

Medical device recalls 2018

Camera System — Class II medical device recall Z-0824-2018

Terminated recall by Trumpf Medical Systems, Inc., reported 2018-03-07, distributed in GA, IL, IN, PA, SC, TX, UT, VA. The radiated emission from the TruVidia" Wireless Receiver is above the normative accepted limits, in which th

Recall numberZ-0824-2018
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmTrumpf Medical Systems, Inc., Charleston, SC, United States
Recall initiated2017-12-15
Classified2018-03-01
Reported by FDA2018-03-07
Terminated2020-04-23
DistributedGA, IL, IN, PA, SC, TX, UT, VA
Countries outside the USAE, AT, AU, BD, BE, CH, CZ, DE, DK, ES, FR, GB, IE, IT, KW, LT, PL, PT, RO, SE, SI, TH, TR
Quantity212 devices
Serial numbers102584474, 102584475, 102584477, 102584478, 102805330, 102813656, 102813657, 102813658, 102813659, 102813660, 102813661, 102813662, 102860647, 102860652, 102863706, 102863707, 102867783, 102867784, 102925734, 102925735, 102925736, 102925737, 102925738, 102925739, 102925740 and 215 more
Model numbers1940747

Product

Camera System, Television, Surgical without audio Intended for transmission and documentation of surgical interventions for the purpose of archiving and teaching.

Reason for recall

The radiated emission from the TruVidia" Wireless Receiver is above the normative accepted limits, in which the radiated emission may interact with other devices in the operating room and cause them not to function properly.