Data Foundry

Medical device recalls 2016

Evita Infinity V500 Ventilator — Class I medical device recall Z-0827-2016

Terminated recall by Draeger Medical, Inc., reported 2016-03-02, distributed nationwide. The battery capacity of the optional PS500 power supply unit for the Infinity (ACS) Workstation Critical and N

Recall numberZ-0827-2016
Product typeMedical device
ClassificationClass I
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmDraeger Medical, Inc., Telford, PA, United States
Recall initiated2016-02-04
Classified2016-02-25
Reported by FDA2016-03-02
Terminated2020-04-01
DistributedNationwide (US)
Quantity2,553
Reason classessoftware
Model numbers8416400, 8417400

Product

Evita Infinity V500 Ventilator; Babylog Infinity VN500 Ventilator. Ventilation.

Reason for recall

The battery capacity of the optional PS500 power supply unit for the Infinity (ACS) Workstation Critical and Neonatal Care was rapidly and unexpectedly reduced, despite using the most recent power supply firmware version 1.50 released in November 2015 (recall Z-0436-2016). The batteries were prematurely discharged, even though an adequate battery charge status was displayed on the device.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0827-2016. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.