Data Foundry

Medical device recalls 2025

nuwellis AquaFlexFlow UF 500 Plus — Class I medical device recall Z-0831-2025

Completed recall by Nuwellis Inc, reported 2025-01-22, distributed nationwide. The AquaFlexFlow UF 500 Plus extracorporeal blood circuit used with the Aquadex SmartFlow and FlexFlow Systems

Recall numberZ-0831-2025
Product typeMedical device
ClassificationClass I
StatusCompleted
Voluntary or mandatedVoluntary
Recalling firmNuwellis Inc, Eden Prairie, MN, United States
Recall initiated2024-12-11
Classified2025-01-16
Reported by FDA2025-01-22
DistributedNationwide (US)
Countries outside the USHK
Quantity845 units
Lot numbers22697, 22698, 22699, 22700, 22701, 22702, 22703, 22704, 22705, 22706, 22707, 22708, 22721, 22722, 22723, 22724, 22733, 22734, 22735, 22736, 22737, 22738, 22739, 22740
Model numbers114156, A06163

Product

nuwellis AquaFlexFlow UF 500 Plus, REF 114156, Catalog Number A06163, extracorporeal blood circuit which is used with the Aquadex SmartFlow or Aquadex FlexFlow System

Reason for recall

The AquaFlexFlow UF 500 Plus extracorporeal blood circuit used with the Aquadex SmartFlow and FlexFlow Systems may indicate "Ultrafiltrate Weight Mismatch" or Excessive Weight Mismatch Alarms while in use. If not addressed, this failure could result in excess fluid removal from a patient leading to Acute Volume Depletion. This failure is especially serious when the Aquadex System is being used on pediatric patients.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0831-2025. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.