adaPT insight — Class II medical device recall Z-0836-2015
Terminated recall by Ion Beam Applications S.A., reported 2014-12-31. Two issues were identified: 1. In the release of adaPT insight v1.3.2., the kV/kV Single Source acquisition wo
| Recall number | Z-0836-2015 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Ion Beam Applications S.A., Louvain La Neuve, Belgium |
| Recall initiated | 2014-11-07 |
| Classified | 2014-12-19 |
| Reported by FDA | 2014-12-31 |
| Terminated | 2015-06-29 |
| Quantity | 1 |
Product
adaPT insight (12C), v1.3.2, PAT.109 (US), Treatment room 4: I2C is used with a charged particle or photon radiation therapy system for localization of the patient position with respect to the therapy equipment and to provide correction feedback to the radiation therapy device.
Reason for recall
Two issues were identified: 1. In the release of adaPT insight v1.3.2., the kV/kV Single Source acquisition workflow was linked to a wrong geometrical calibration file, resulting in an alignment offset for this specific mode of acquisition. 2. A gantry collision incident required the IBA service team to verify the gantry alignment and protons I x-ray beams colinearity, du