Ysiosystem with VC10 software version — Class II medical device recall Z-0841-2022

Ongoing recall by Siemens Medical Solutions USA, Inc, reported 2022-04-06, distributed nationwide. Siemens Healthineers has identified a software error in previous software version VB10 for the Ysio fully auto

Recall numberZ-0841-2022
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmSiemens Medical Solutions USA, Inc, Malvern, PA, United States
Recall initiated2022-01-31
Classified2022-03-30
Reported by FDA2022-04-06
DistributedNationwide (US)
Quantity70 systems
Reason classessoftware
Serial numbers20105, 20113, 20115, 20118, 21004, 21005, 21108, 21114, 21174, 21194, 21209, 22029, 22054, 22058, 22071, 22076, 22092, 22104, 22119, 22126, 22128, 22135, 22136, 22146, 22163 and 45 more
Model numbers10281013

Product

Ysiosystem with VC10 software version, Model Number 10281013. Radiographic system

Reason for recall

Siemens Healthineers has identified a software error in previous software version VB10 for the Ysio fully automated system. The unit may process incorrect internal values for the position of its ceiling tube stand. If the user activates a longitudinal or transversal movement at the time of the error, the system controls may initiate a correction of the tube stand position, resulting in an unintended movement. If the described error occurs, a potential for collision risk for objects or persons in the vicinity of the moving stand exists. This may result in minor to serious injuries to persons.