Data Foundry

Medical device recalls 2019

Aplicare Povidone Iodine Prep Pads packaged within Portex Pro-Vent… — Class II medical device recall Z-0842-2019

Terminated recall by Smiths Medical ASD Inc., reported 2019-02-27, distributed nationwide. The povidone iodine prep pads packaged with the arterial blood sampling kits will not maintain the iodine pote

Recall numberZ-0842-2019
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmSmiths Medical ASD Inc., Minneapolis, MN, United States
Recall initiated2018-06-22
Classified2019-02-15
Reported by FDA2019-02-27
Terminated2020-10-05
DistributedNationwide (US)
Countries outside the USAE, BS, CA, CR, LV, SA
Quantity28,578,939 kits
Reason classeslabeling, potency
Model numbers4589P-1, 4591EG-1, 4591P-1, 4599P-1, 4611LH, 4611P-1, 4620EG-1, 4620P-1, 4640LH, 4640P-1, 4649P-1, 4660EG-1, 4660P-1, 4699LH, 4699P-1, G1325, G1412, G1481, G1509, G1548, G1581, G1586, G1602, G1627, G1632 and 7 more

Product

Aplicare Povidone Iodine Prep Pads packaged within Portex Pro-Vent Arterial Blood Sampling Kit with Dry Lithium Heparin for Gases and Electrolytes, models 4589P-1, 4591EG-1, 4591P-1, 4599P-1, 4611LH, 4611P-1, 4620EG-1, 4620P-1, 4640LH, 4640P-1, 4649P-1, 4660EG-1, 4660P-1, 4699LH, 4699P-1, G1325, G1412, G1481, G1509, G1548, G1581, G1586, G1602, G1627, G1632, G1634, G1671, G1675, G1716, G1724, G1773, and G1807. The firm name on the label is Smiths Medical ASD, Inc., Keene, NH.

Reason for recall

The povidone iodine prep pads packaged with the arterial blood sampling kits will not maintain the iodine potency through its labeled 36-month expiration dating.