Data Foundry

Medical device recalls 2020

Instrumentarium Dental Express SOLID STATE X-RAY IMAGER — Class II medical device recall Z-0842-2020

Terminated recall by Palodex Group Oy, reported 2020-01-29, distributed in PA, WI. Slow degradation of plastic insulation used in the PSU DC plug of the DIGORA Optime, Scan eXam, Express, GXPS-

Recall numberZ-0842-2020
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmPalodex Group Oy, Tuusula, Finland
Recall initiated2020-01-07
Classified2020-01-23
Reported by FDA2020-01-29
Terminated2024-04-15
DistributedPA, WI
Countries outside the USAE, AR, AT, AU, BE, BR, CA, CH, CL, CN, CR, CZ, DE, DK, DZ, EC, ES, FI, FR, GB, GR, HU, IL, IN, IT, JP, KR, KW, MT, MY, NL, NO, NZ, OM, PK, PL, QA, RO, RS, SA, SE, SG, SI, TH, TR, TW, VN
Quantity4,563 total
Reason classesspecification failure, device malfunction

Product

Instrumentarium Dental Express SOLID STATE X-RAY IMAGER (FLAT PANEL/DIGITAL IMAGER)

Reason for recall

Slow degradation of plastic insulation used in the PSU DC plug of the DIGORA Optime, Scan eXam, Express, GXPS-500 systems may result in short circuiting which may cause heating and melting. The melting may generate smoke and eventually cause the PSU to fail.