Data Foundry

Medical device recalls 2014

Simplexa Flu A/B & RSV Direct assay — Class II medical device recall Z-0846-2014

Terminated recall by Focus Diagnostics Inc, reported 2014-02-05, distributed nationwide. Focus Diagnostics is initiating an urgent safety notice correction for Simplexa Flu A/B & RSV Direct assay (MO

Recall numberZ-0846-2014
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmFocus Diagnostics Inc, Cypress, CA, United States
Recall initiated2014-01-10
Classified2014-01-28
Reported by FDA2014-02-05
Terminated2014-10-15
DistributedNationwide (US)
Countries outside the USCA, IL, SE, SI
Quantity1,052 kits
Reason classespotency
Lot numbers24493, 24495, 24535, 24536
Model numbersMOL2650

Product

Simplexa Flu A/B & RSV Direct assay, Model MOL2650. The Focus Diagnostics Simplexa Flu A/B & RSV Direct assay is intended for use on the 3M Integrated Cycler instrument for the in vitro qualitative detection and differentiation of influenza A virus, influenza B virus, and respiratory syncytial virus (RSV) RNA in nasopharyngeal swabs (NPS) from human patients with signs and symptoms of respiratory tract infection in conjunction with clinical and epidemiological risk factors. This test is intended for use as an aid in the differential diagnosis of influenza A, influenza B, and RSV viral infections in humans and is not intended to detect influenza C.

Reason for recall

Focus Diagnostics is initiating an urgent safety notice correction for Simplexa Flu A/B & RSV Direct assay (MOL2650) because Focus Diagnostics has received some customer complaints of Simplexa Flu A/B & RSV lots due to sporadic false positive RSV signals, which may result in a higher RSV false positive rate.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0846-2014. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.