Philips Network Firewall — Class II medical device recall Z-0850-2018
Terminated recall by Philips Electronics North America Corporation, reported 2018-03-14, distributed nationwide. Firewall installed with Philips IntelliVue Information Center iX or Information Center Classic may have a defe
| Recall number | Z-0850-2018 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Philips Electronics North America Corporation, Andover, MA, United States |
| Recall initiated | 2017-10-05 |
| Classified | 2018-03-02 |
| Reported by FDA | 2018-03-14 |
| Terminated | 2019-07-01 |
| Distributed | Nationwide (US) |
| Quantity | 332 units |
| Serial numbers | JMX194840WA, JMX2014419Y, JMX201441A4, JMX201441AB, JMX2014Z155, JMX2014Z15C, JMX2014Z15L, JMX20184033, JMX2018Z009, JMX2022Y15F, JMX2022Y181, JMX2022Y194, JMX2022Y1A7, JMX2022Y1U4, JMX2030Y2A8, JMX2030Y2AY, JMX2030Y2B1, JMX2030Y2B6, JMX2030Y2B8, JMX2030Y2BN, JMX2030Y2BT, JMX2030Y2CR, JMX2030Y2D9, JMX2030Y2DM, JMX2030Y2DY and 22 more |
| Model numbers | 867098 |
Product
Philips Network Firewall (Cisco ASA 5506), Model # 867098
Reason for recall
Firewall installed with Philips IntelliVue Information Center iX or Information Center Classic may have a defective component, which may result in loss of connection to the Information Center iX. The defect involves the clock signal component within the firewall. This component has a high probability of failing in appliances that have been running for greater than 18 months. If the clock signal component were to fail, the firewall will stop functioning, will not boot, and is not recoverable. This failure will result in loss of communication between devices that are separated by the firewall, which may cause the Information Center to reboot.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0850-2018. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.