User Defined Method Flex Assignment/Siemens Dimension Vista 500 or… — Class II medical device recall Z-0851-2014
Terminated recall by Siemens Healthcare Diagnostics, Inc., reported 2014-02-05, distributed nationwide. When utilizing the Routine Inventory screen to enter a User Defined Method (EMPTY) Flex the system may assign
| Recall number | Z-0851-2014 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Siemens Healthcare Diagnostics, Inc., Newark, DE, United States |
| Recall initiated | 2013-10-07 |
| Classified | 2014-01-28 |
| Reported by FDA | 2014-02-05 |
| Terminated | 2014-08-27 |
| Distributed | Nationwide (US) |
| Quantity | 2,038 |
Product
User Defined Method Flex Assignment/Siemens Dimension Vista 500 or Dimension Vista 1500 System with software version 3.5.1 or lower User Defined Method Flex Assignment/Siemens Dimension Vista 500 or Dimension Vista 1500 System is an in vitro diagnostic device intended to duplicate manual analytical procedures such as pipetting, mixing, heating, and measuring spectral intensities to determine a variety of analytes in human body fluids.
Reason for recall
When utilizing the Routine Inventory screen to enter a User Defined Method (EMPTY) Flex the system may assign the User Defined Method Flex to a different Flex that is currently in inventory on the system, and then use the incorrect Flex Cartridge to process the user defined method.