Data Foundry

Medical device recalls 2014

GE Healthcare Innova IGS 630 — Class II medical device recall Z-0853-2014

Terminated recall by GE Healthcare, LLC, reported 2014-02-05, distributed nationwide. GE Healthcare has recently become aware of a potential safety issue with respect to IGS 630 Imaging Systems. W

Recall numberZ-0853-2014
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmGE Healthcare, LLC, Waukesha, WI, United States
Recall initiated2013-11-15
Classified2014-01-28
Reported by FDA2014-02-05
Terminated2014-03-28
DistributedNationwide (US)
Countries outside the USCZ, FR, JP, KR, SA, TR

Product

GE Healthcare Innova IGS 630. Biplane Cardiovascular and Interventional Imaging System. Product Usage: The angiographic X-ray systems are indicated for use for patients from newborn to geriatric in generating fluoroscopic and rotational images of human anatomy for cardiovascular, vascular and non-vascular, diagnostic and interventional procedures. Additionally, with the OR table, the angiographic X-ray systems are indicated for use in generating fluoroscopic and rotational images of human anatomy for image-guided surgical procedures. The OR table is suitable for interventional and surgical procedures. The Cathiab Frontiers solutions are indicated for use in conjunction with single plane and biplane GE angiographic X-ray systems and imaging / data medical devices used in interventional and surgical cathlab environments and cleared for commercial distribution. The Cathiab Frontiers solutions are integrated GE angiographic X-ray and imaging / data medical devices that simplify the end-to-end clinical workflow in the cathlab by implementing: 1) Communication protocols for exchanging and automatically synchronizing patient, exam, system, and image information between the angiographic X-ray systems and the imaging / data medical devices. 2) Communication protocols for the control of imaging / data medical device functions from the angiographic X-ray systems user interface. 3) Interfaces for displaying the imaging / data medical device output on the monitor display solutions of the GE angiographic X-ray systems.

Reason for recall

GE Healthcare has recently become aware of a potential safety issue with respect to IGS 630 Imaging Systems. While performing fluoroscopy on a biplane fluoroscopy unit, there is a potential for loss of the x-ray imaging function when the user changes the size of the lateral FOV (Field of View) and releases the biplane footswitch pedal simultaneously. This could lead to a system lock up requiring

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0853-2014. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.