GE Healthcare Innova IGS 630 — Class II medical device recall Z-0853-2014
Terminated recall by GE Healthcare, LLC, reported 2014-02-05, distributed nationwide. GE Healthcare has recently become aware of a potential safety issue with respect to IGS 630 Imaging Systems. W
| Recall number | Z-0853-2014 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | GE Healthcare, LLC, Waukesha, WI, United States |
| Recall initiated | 2013-11-15 |
| Classified | 2014-01-28 |
| Reported by FDA | 2014-02-05 |
| Terminated | 2014-03-28 |
| Distributed | Nationwide (US) |
| Countries outside the US | CZ, FR, JP, KR, SA, TR |
Product
GE Healthcare Innova IGS 630. Biplane Cardiovascular and Interventional Imaging System. Product Usage: The angiographic X-ray systems are indicated for use for patients from newborn to geriatric in generating fluoroscopic and rotational images of human anatomy for cardiovascular, vascular and non-vascular, diagnostic and interventional procedures. Additionally, with the OR table, the angiographic X-ray systems are indicated for use in generating fluoroscopic and rotational images of human anatomy for image-guided surgical procedures. The OR table is suitable for interventional and surgical procedures. The Cathiab Frontiers solutions are indicated for use in conjunction with single plane and biplane GE angiographic X-ray systems and imaging / data medical devices used in interventional and surgical cathlab environments and cleared for commercial distribution. The Cathiab Frontiers solutions are integrated GE angiographic X-ray and imaging / data medical devices that simplify the end-to-end clinical workflow in the cathlab by implementing: 1) Communication protocols for exchanging and automatically synchronizing patient, exam, system, and image information between the angiographic X-ray systems and the imaging / data medical devices. 2) Communication protocols for the control of imaging / data medical device functions from the angiographic X-ray systems user interface. 3) Interfaces for displaying the imaging / data medical device output on the monitor display solutions of the GE angiographic X-ray systems.
Reason for recall
GE Healthcare has recently become aware of a potential safety issue with respect to IGS 630 Imaging Systems. While performing fluoroscopy on a biplane fluoroscopy unit, there is a potential for loss of the x-ray imaging function when the user changes the size of the lateral FOV (Field of View) and releases the biplane footswitch pedal simultaneously. This could lead to a system lock up requiring