smiths medical Medfusion Model 4000 syringe pump — Class I medical device recall Z-0864-2024
Ongoing recall by Smiths Medical ASD Inc., reported 2024-02-07, distributed nationwide. Medfusion Model 4000 syringe infusion pumps, with the following software versions, may have the following issu
| Recall number | Z-0864-2024 |
|---|---|
| Product type | Medical device |
| Classification | Class I |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Smiths Medical ASD Inc., Minneapolis, MN, United States |
| Recall initiated | 2023-12-19 |
| Classified | 2024-02-01 |
| Reported by FDA | 2024-02-07 |
| Distributed | Nationwide (US) |
| Countries outside the US | AU, BM, CA, DE, GB, LB, MY, SA, SG |
| Quantity | 60,146 pumps |
| Reason classes | software |
| Model numbers | 4000 |
Product
smiths medical Medfusion Model 4000 syringe pump
Reason for recall
Medfusion Model 4000 syringe infusion pumps, with the following software versions, may have the following issues: 1. Delivery During Motor Not Running High Priority Alarm (v1.0.0, v1.1.0, v1.1.1, v1.1.2), 2. Re-administered Loading Dose (v1.0.0, v1.1.0, v1.1.1, v1.1.2), 3. Incorrect Critical Data Failure Alarm, 4. Interruption of Bolus or Loading Dose Delivery (v1.0.0, v1.1.0, v1.1.1, v1.1.2), 5. Incorrect Total Bolus/Loading Dose Displayed (v1.1.0, v1.1.1, v1.1.2), 6. Volume Limit Before Bolus/Loading Dose Complete (v1.1.2), 7. Drug Library Lower Limit Displayed Incorrectly (v1.6.0, v1.6.1), 8. Depleted Battery Alarm (All versions previous to v1.6.5), 9. Loss of Wireless Connectivity (v1.5.0, v1.5.1, v1.6.0, v1.6.1, v1.6.4), 10. PharmGuard Server Password (v2.3, v2.4, v2.5). Smiths Medical corrected all issues included in this notification in previous software updates and the corrections were carried forward into all subsequent software releases. Please ensure you have the most recent Medfusion software installed on your pumps.