Reliance Vision Single-Chamber Washer/Disinfector — Class II medical device recall Z-0876-2021
Terminated recall by Steris Corporation, reported 2021-01-27, distributed nationwide. The electrical contactor component present in the drying chamber of the Reliance Vision Single-Chamber Washer/
| Recall number | Z-0876-2021 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Steris Corporation, Mentor, OH, United States |
| Recall initiated | 2020-12-11 |
| Classified | 2021-01-15 |
| Reported by FDA | 2021-01-27 |
| Terminated | 2022-06-21 |
| Distributed | Nationwide (US) |
| Countries outside the US | AE, AU, BE, BG, BR, CA, CH, CL, CN, DE, EC, ES, FR, GB, HK, IE, IL, IT, JP, KR, MX, MY, NZ, PA, PT, SA, SG, TW, VE |
| Quantity | 3,405 |
| Reason classes | device malfunction |
| Serial numbers | 3600311010, 3600312001, 3600312004, 3600312017, 3600312018, 3600317007, 3600317008, 3600317015, 3600317016, 3600317017, 3600317018, 3600317019, 3600318002, 3600318004, 3600410012, 3600416017, 3600416018, 3600416019, 3600416020, 3600416022, 3600416023, 3600419027, 3600419028, 3600419029, 3600510010 and 475 more |
Product
Reliance Vision Single-Chamber Washer/Disinfector - Product Usage: intended for use in the cleaning and intermediate level disinfection of soiled reusable utensils, trays, glassware, bedpans and urinals, rubber and plastic goods, simple hard-surfaced rigid surgical instruments, such as forceps and clamps, theatre shoes, and other similar and related articles found in healthcare facilities.
Reason for recall
The electrical contactor component present in the drying chamber of the Reliance Vision Single-Chamber Washer/Disinfector may malfunction, potentially leading to the heating elements in the drying chamber to overheat, eventually resulting in smoke and/or fire.