Data Foundry

Medical device recalls 2013

BMC RF Cannula Curved Sharp RadiOpaque — Class II medical device recall Z-0879-2013

Terminated recall by Baylis Medical Company Inc., reported 2013-03-06. A report of an event was received indicating that upon opening a box containing 10 sterile cannula by a nurse,

Recall numberZ-0879-2013
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmBaylis Medical Company Inc., Mississauga, Ontario, Canada
Recall initiated2012-10-04
Classified2013-02-26
Reported by FDA2013-03-06
Terminated2013-05-23
Countries outside the USAU
Quantity1,780
Reason classessterility
Model numbers16-145-10CS

Product

BMC RF Cannula Curved Sharp RadiOpaque, Non-Pyrogenic, Active Tip: 10mm, Gauge 16 Ga, Length: 100mm. The BMC RF Cannula is sealed in a Tyvek pouch (primary packaging) and then placed into a white shelf-box (secondary packaging) with the instructions for use. A predetermined number of white shelf-boxes are then placed into shipping cartons (tertiary packaging). The BMC RF Cannula is intended for use in radiofrequency heat lesion procedures for relief of pain.

Reason for recall

A report of an event was received indicating that upon opening a box containing 10 sterile cannula by a nurse, the pouch package of 5 of the 10 individually packaged cannula had not been sealed along one edge. Thus the pouch was open and the device could fall out of the pouch and on to the floor when retrieved from the shelf box. The cannulas were not used on any patients, as the unsealed packages