Data Foundry

Medical device recalls 2015

Alere INRatio PT/INR Test Strips — Class I medical device recall Z-0880-2015

Terminated recall by Alere San Diego, Inc., reported 2015-01-14, distributed nationwide. In certain cases, an INRatio PT/INR Monitor system may provide an INR result that is significantly lower than

Recall numberZ-0880-2015
Product typeMedical device
ClassificationClass I
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmAlere San Diego, Inc., San Diego, CA, United States
Recall initiated2014-12-05
Classified2015-01-03
Reported by FDA2015-01-14
Terminated2017-09-18
DistributedNationwide (US)
Countries outside the USCA
Reason classeslabeling
Model numbers0100071, 0100139

Product

Alere INRatio PT/INR Test Strips, Alere INRatio PT/INR System Professional - 0100071 Alere INRatio PT/INR Test Strips, Box of 12 - 0100139 Alere INRatio PT/INR Test Strips, Box of 48 The test strips are packaged in individually pouched test strips in a labeled box with 12 or 48 test strips per box. In vitro diagnostic system that provides a quantitative prothrombin time result, expressed in seconds and as an International normalized ratio (INR).

Reason for recall

In certain cases, an INRatio PT/INR Monitor system may provide an INR result that is significantly lower than a result obtained using a reference INR system (laboratory method). This issue can arise if the patient has certain medical conditions. It can also occur if the instructions in the labeling for performing the test are not followed.