Data Foundry

Medical device recalls 2021

Abbott i-STAT CG8+ cartridges — Class II medical device recall Z-0880-2021

Terminated recall by Abbott Point Of Care Inc., reported 2021-01-27, distributed nationwide. Ionized calcium test on some i-STAT CG8+ and EG7+ cartridges may exhibit higher than expected reported results

Recall numberZ-0880-2021
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmAbbott Point Of Care Inc., Princeton, NJ, United States
Recall initiated2020-12-22
Classified2021-01-19
Reported by FDA2021-01-27
Terminated2023-09-01
DistributedNationwide (US)

Product

Abbott i-STAT CG8+ cartridges - IVD Intended for quantitative measurement of sodium, potassium, ionized calcium, glucose, pH, pCO2, pO2 and hematocrit in venous, arterial or capillary whole blood List Number: 03P88-25

Reason for recall

Ionized calcium test on some i-STAT CG8+ and EG7+ cartridges may exhibit higher than expected reported results when stored at room temperature (18 to 30C) for periods of time in excess of 3 days for lot numbers *20100 to *20339 and 7 days for lot numbers *20340 and above Where * = lot letter A, K, L, M, N, W or Y.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0880-2021. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.