Data Foundry

Medical device recalls 2013

Philips Medical Systems — Class II medical device recall Z-0881-2013

Terminated recall by Philips Medical Systems (Cleveland) Inc, reported 2013-03-06, distributed in AK, AZ, CA, CO, CT, DC, DE, FL…. Software anomaly. Philips determined that the TumorLOC software program for the referenced PET/CT X-Ray scanni

Recall numberZ-0881-2013
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmPhilips Medical Systems (Cleveland) Inc, Cleveland, OH, United States
Recall initiated2012-10-24
Classified2013-02-26
Reported by FDA2013-03-06
Terminated2015-07-23
DistributedAK, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, MA, MD, MI, MN, MO, MS, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV
Countries outside the USCA
Quantity39 units
Reason classeslabeling, software
Model numbers728260

Product

Philips Medical Systems, Extended Brilliance Workspace (EBW) X-Ray Scanners, Model #728260. A component of the Computed Tomography X-Ray System.

Reason for recall

Software anomaly. Philips determined that the TumorLOC software program for the referenced PET/CT X-Ray scanning systems has the capability to generate Intensity Projection data sets from respiratory gated data. A circumstance was identified by the firm whereby these generated datasets are 'flipped' and incorrectly labeled left-to-right when the are saved to disk. If the flipped Intensity Project

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0881-2013. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.