Data Foundry

Medical device recalls 2025

Incisive CT — Class II medical device recall Z-0881-2026

Ongoing recall by Philips North America Llc, reported 2025-12-10, distributed nationwide. If the screws of tube heat exchanger on the rotating scanner are not tightened well after replacing. This comp

Recall numberZ-0881-2026
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmPhilips North America Llc, Cambridge, MA, United States
Recall initiated2025-10-29
Classified2025-12-03
Reported by FDA2025-12-10
DistributedNationwide (US)
Countries outside the USAR, AT, AU, BD, BE, BG, BH, BO, BR, BS, CA, CH, CL, CN, CO, CR, CY, CZ, DE, DK, DZ, EC, EG, ES, FR, GB, GR, HK, HR, HU, ID, IL, IN, IQ, IR, IT, JO, JP, KE, KR, KW, LB, LT, LV, MA, MX, MY, NG, NI, NL, NO, NP, NZ, OM, PA, PE, PH, PK, PL, PT, PY, QA, RO, RS, SA, SE, SK, SV, TH, TN, TT, TW, UA, VE, VN, ZA
Quantity2,341 units
Reason classesdevice malfunction
UPC / GTIN / UDI-DI00884838085015, 00884838105508, 55400406110738, 55400406111117, 55400406111438, 55400406111629, 55400406111636
Serial numbers33001, 33002, 33003, 33005, 33006, 33007, 33008, 33009, 33010, 33011, 33012, 33013, 33015, 33016, 33017, 33018, 33019, 33020, 33021, 33022, 33023, 33024, 33025, 33026, 33027 and 475 more
Model numbers728143, 728144

Product

Incisive CT; Software Version Number: 4.5, 5.0, 5.1;

Reason for recall

If the screws of tube heat exchanger on the rotating scanner are not tightened well after replacing. This component may become detached and make contact with other components located within the Incisive CT systems during rotation. Other components could be damaged due to the contact with the tube heat exchanger.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0881-2026. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.