Data Foundry

Medical device recalls 2015

INRatio 2 Monitors — Class I medical device recall Z-0882-2015

Terminated recall by Alere San Diego, Inc., reported 2015-01-14, distributed nationwide. In certain cases, an INRatio PT/INR Monitor system may provide an INR result that is significantly lower than

Recall numberZ-0882-2015
Product typeMedical device
ClassificationClass I
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmAlere San Diego, Inc., San Diego, CA, United States
Recall initiated2014-12-05
Classified2015-01-03
Reported by FDA2015-01-14
Terminated2017-09-18
DistributedNationwide (US)
Countries outside the USCA
Quantity231,967 test
Reason classeslabeling
Model numbers0200431, 0200432, 0200433, 0200457, 55112, 55128A, 55131

Product

INRatio 2 Monitors - 0200431 Alere INRatio 2 PT/INR Professional Testing System - 0200432 Alere INRatio 2 PT/INR Home Monitoring System - 0200433 Alere INRatio 2 PT/INR Home Monitoring System - 0200457 Alere INRatio 2 Replacement Monitor - 55112 Alere INRatio 2 Replacement Monitor - 55131 Alere INRatio 2 Replacement Monitor - 55128A Alere INRatio 2 PT/INR Professional Monitoring System The test systems contain individually packaged components needed to use the product, such as a monitor, user guide, quick reference guide training DVD or CD, Lancets (home monitoring kits contain an Autolet, Unilet, gauze pads, and alcohol wipes), power supply (home monitoring kits contain AA batteries), results log book and a warranty card. None of the components, except for the lancets, are sterile. Intended for in vitro diagnostic use for quantitative determination of international normalized ratio (INR) in fresh capillary whole blood to monitor the effect of warfarin on clotting time by health care professionals and home users.

Reason for recall

In certain cases, an INRatio PT/INR Monitor system may provide an INR result that is significantly lower than a result obtained using a reference INR system (laboratory method). This issue can arise if the patient has certain medical conditions. It can also occur if the instructions in the labeling for performing the test are not followed.