Philips Respironics V60 Ventilator Part Number 1053613 — Class II medical device recall Z-0885-2021
Terminated recall by Respironics California, LLC, reported 2021-02-03, distributed nationwide. Ventilator may experience potential issues with the backup battery failures: 1) if the battery is allowed to o
| Recall number | Z-0885-2021 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Respironics California, LLC, Carlsbad, CA, United States |
| Recall initiated | 2020-10-30 |
| Classified | 2021-01-22 |
| Reported by FDA | 2021-02-03 |
| Terminated | 2024-09-19 |
| Distributed | Nationwide (US) |
| Countries outside the US | AE, AF, AR, AT, AU, BD, BE, BG, BH, BO, BR, CA, CH, CL, CN, CO, CY, CZ, DE, DK, DZ, EC, EG, ES, FI, FR, GB, GH, GR, GT, HK, HR, HU, ID, IE, IL, IN, IR, IS, IT, JO, JP, KE, KR, LB, LU, LV, MA, MT, MX, MY, NG, NL, NO, NP, NZ, OM, PA, PE, PH, PK, PL, PT, PY, QA, RO, SA, SE, SG, SI, TH, TR, TW, UA, VE, VN, ZA |
| Quantity | 2,144 systems |
| Serial numbers | 100026086, 100026094, 100026121, 100026664, 100026671, 100026672, 100026678, 100026686, 100026755, 100026760, 100028670, 100029057, 100029058, 100029066, 100029068, 100029074, 100029075, 100029077, 100029078, 100029222, 100029240, 100029335, 100029457, 100029459, 100029460 and 475 more |
| Model numbers | 1053613, V60 |
Product
Philips Respironics V60 Ventilator Part Number 1053613 - Product Usage: intended to be used only with various combinations of Respironics-recommended patient circuits, interfaces (masks), humidifiers, and other accessories.
Reason for recall
Ventilator may experience potential issues with the backup battery failures: 1) if the battery is allowed to over discharge, or 2) if the ventilator is stored for an extended period of time without being plugged into AC power.