Data Foundry

Medical device recalls 2015

ACCU-CHEK Tender I 17/60 10 pieces Product Usage: Intended for the… — Class II medical device recall Z-0886-2015

Terminated recall by Roche Diagnostics Operations, Inc., reported 2015-01-07, distributed nationwide. Roche Diabetes Care was informed by UnoMedical, the Manufacturer of ACCU-CHEK Tender I and ACCU-CHEK Tender II

Recall numberZ-0886-2015
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmRoche Diagnostics Operations, Inc., Indianapolis, IN, United States
Recall initiated2014-11-20
Classified2014-12-30
Reported by FDA2015-01-07
Terminated2018-01-10
DistributedNationwide (US)
Quantity13,041 boxes
Reason classesdevice malfunction
Lot numbers5040425, 5044367, 5051305, 5053294, 5061932
Model numbers04541430001, 5040425, 5044367, 5051305, 5053294, 5061932

Product

ACCU-CHEK Tender I 17/60 10 pieces Product Usage: Intended for the infusion and/or injection of fluids into the body below the surface of the skin. The indwelling catheter can be inserted independently from the infusion catheter and can be accessed for the injections through the injection port. The indwelling catheter can also be securely attached to the infusion catheter by means of a proprietary click-lock connector for the infusion of drugs subcutaneously. The infusion set can be detached from the indwelling catheter, and the catheter capped to allow freedom from the infusion set and pump for showers, athletics or other activities.

Reason for recall

Roche Diabetes Care was informed by UnoMedical, the Manufacturer of ACCU-CHEK Tender I and ACCU-CHEK Tender II infusion sets, about potential for the the tubing of the infusion set to become detached at the connect/disconnect location on the ACCU-CHEK Tender I and ACCU-CHEK Tender II infusion sets. If tubing detachment occurs, insulin delivery is interrupted and the pump will not alarm to notify