Data Foundry

Medical device recalls 2013

Firebird Spinal Fixation System — Class II medical device recall Z-0896-2013

Terminated recall by Orthofix, Inc, reported 2013-03-06, distributed nationwide. There is a possibility that the Set Screw (p/n 90-2923) sub-component of the Parallel Rod Connector Assembly (

Recall numberZ-0896-2013
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmOrthofix, Inc, Lewisville, TX, United States
Recall initiated2013-02-01
Classified2013-02-28
Reported by FDA2013-03-06
Terminated2013-07-01
DistributedNationwide (US)
Countries outside the USAU, DE, ES, IL, LB, SE
Reason classespackaging
Model numbers52-6700, 52-6800

Product

Firebird Spinal Fixation System, Parallel Rod Connector (52-6800) and Axial Rod Connector (52-6700). Product Usage: The connector assemblies allow for rod components to be connected side-to-side, or end-to-end and eliminate the need of rebuilding the entire system with a longer rod in order to include the new vertebral levels.

Reason for recall

There is a possibility that the Set Screw (p/n 90-2923) sub-component of the Parallel Rod Connector Assembly (p/n 52-6800) & Axial Rod Connector (p/n 52-6700) may crack during final tightening. Orthofix Inc received six (6) customer complaints related to set screw cracking between November 15, 2010 and December 2012.