LIPOSORBER System — Class II medical device recall Z-0900-2022
Terminated recall by Kaneka America Corp, reported 2022-04-20, distributed in AL, AZ, CA, CO, CT, DE, FL, GA…. Labeling change to due to the risk of severe anaphylactoid reactions including shock in patients treated with
| Recall number | Z-0900-2022 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Kaneka America Corp, New York, NY, United States |
| Recall initiated | 2022-02-17 |
| Classified | 2022-04-08 |
| Reported by FDA | 2022-04-20 |
| Terminated | 2024-05-06 |
| Distributed | AL, AZ, CA, CO, CT, DE, FL, GA, IL, IN, KS, MA, MD, ME, MI, MN, MO, MS, NC, NE, NH, NJ, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI |
| Reason classes | labeling |
| Model numbers | KP-05, LA-15, MA-03 |
Product
LIPOSORBER System, Model: LA-15, consisting of: Apheresis machine, product# 101445 model: MA-03, and disposables (product#, model) [#101447, KP-05 Sulflux Plasma Separator], [#101473, HDE KP-05 Sulflux Plasma Separator], [#101785, LA-15 (AU) LDL Adsorption Column (Luer Lock Type)], [#101786, NK-M3R(UL) Tubing (Luer Lock Type)], [#101788, HDE LA-15(AU) LDL Adsorption Column (Luer Lock Type)], [#101789, HDE NK-M3R(UL) Tubing (Luer Lock Type)]. Associated labeling: 1) Instructions for Use (6): IFU LIPOSORBER LA-15 LDL Adsorption Columns FH, IFU SULFLUX KP-05 PLASMA SE-PARATOR FH, IFU TUBING SYSTEM FOR PLASMAPHERESIS NK-M3R(UL) FH, IFU LIPOSORBER LA-15 LDL Adsorption Columns FSGS, IFU SULFLUX KP-05 PLASMA SE-PARATOR FSGS, IFU TUBING SYSTEM FOR PLASMAPHERESIS NK-M3R(UL) FSGS, 2) Operators Manuals for the following (2): LIPOSORBER LA-15 SYSTEM OPERATOR's MANUAL FH, LIPOSORBER LA-15 SYSTEM OPERATOR's MANUAL FSGS 3) Patients Guide (1): A PATIENT GUIDE TO THE LIPOSORBER LA-15 SYSTEM FSGS
Reason for recall
Labeling change to due to the risk of severe anaphylactoid reactions including shock in patients treated with the Liposorber LA-15 LDL Adsorption Column under concomitant ACE inhibitor medication treatment.