Brand Name: VITROS Chemistry Products Product Name: VITROS Chemistry… — Class II medical device recall Z-0901-2025

Ongoing recall by QUIDEL ORTHO, reported 2025-01-22, distributed nationwide. Hemoglobin interference may occur at concentrations lower than the claimed 300 mg/dL, and may cause negatively

Recall numberZ-0901-2025
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmQUIDEL ORTHO, Rochester, NY, United States
Recall initiated2024-12-05
Classified2025-01-10
Reported by FDA2025-01-22
DistributedNationwide (US)
Countries outside the USAU, BE, BM, BR, CA, CL, CN, CO, DE, DK, ES, FR, GB, IN, IT, JP, MX, NL, NO, PT, RU, SE, SG
UPC / GTIN / UDI-DI10758750004782
Model numbers8343386
Expiration dates2025-03-01

Product

Brand Name: VITROS Chemistry Products Product Name: VITROS Chemistry Products DGXN Slides Model/Catalog Number: 8343386 Software Version: Not Applicable Product Description: The VITROS DGXN Slide is a multilayered, analytical element coated on a polyester support. Digoxin is a widely prescribed cardiac glycoside indicated in the treatment of congestive heart failure and supraventricular arrhythmias. Digoxin measurements are used to monitor patient compliance and therapy, and to diagnose potential overdose. VITROS Chemistry Products DGXN Slides quantitatively measure digoxin (DGXN) concentration in serum using VITROS 250/350/5,1 FS/4600/XT 3400 Chemistry Systems and the VITROS 5600/XT 7600 Integrated Systems. One sales unit contains 5 cartridges. Each cartridge holds 18 slides (equivalent to one test), for 90 slides per sales unit. Component: No, not a component

Reason for recall

Hemoglobin interference may occur at concentrations lower than the claimed 300 mg/dL, and may cause negatively biased digoxin results.