Brand Name: VITROS Chemistry Products Product Name: VITROS Chemistry… — Class II medical device recall Z-0902-2025
Ongoing recall by QUIDEL ORTHO, reported 2025-01-22, distributed nationwide. Hemoglobin interference may occur at concentrations lower than the claimed 300 mg/dL, and may cause negatively
| Recall number | Z-0902-2025 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | QUIDEL ORTHO, Rochester, NY, United States |
| Recall initiated | 2024-12-05 |
| Classified | 2025-01-10 |
| Reported by FDA | 2025-01-22 |
| Distributed | Nationwide (US) |
| Countries outside the US | AU, BE, BM, BR, CA, CL, CN, CO, DE, DK, ES, FR, GB, IN, IT, JP, MX, NL, NO, PT, RU, SE, SG |
| UPC / GTIN / UDI-DI | 10758750004782 |
| Model numbers | 8343386 |
| Expiration dates | 2025-04-01 |
Product
Brand Name: VITROS Chemistry Products Product Name: VITROS Chemistry Products DGXN Slides Model/Catalog Number: 8343386 Software Version: Not Applicable Product Description: The VITROS DGXN Slide is a multilayered, analytical element coated on a polyester support. Digoxin is a widely prescribed cardiac glycoside indicated in the treatment of congestive heart failure and supraventricular arrhythmias. Digoxin measurements are used to monitor patient compliance and therapy, and to diagnose potential overdose. VITROS Chemistry Products DGXN Slides quantitatively measure digoxin (DGXN) concentration in serum using VITROS 250/350/5,1 FS/4600/XT 3400 Chemistry Systems and the VITROS 5600/XT 7600 Integrated Systems. One sales unit contains 5 cartridges. Each cartridge holds 18 slides (equivalent to one test), for 90 slides per sales unit. Component: No, not a component
Reason for recall
Hemoglobin interference may occur at concentrations lower than the claimed 300 mg/dL, and may cause negatively biased digoxin results.