Philips Respironics V60 Ventilator Part Number 1076709 — Class II medical device recall Z-0908-2021

Ongoing recall by Respironics California, LLC, reported 2021-02-03, distributed nationwide. Potential issues with ventilator navigation ring interface and liquid ingress: 1) The navigation ring may beco

Recall numberZ-0908-2021
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmRespironics California, LLC, Carlsbad, CA, United States
Recall initiated2020-11-02
Classified2021-01-22
Reported by FDA2021-02-03
DistributedNationwide (US)
Countries outside the USAE, AF, AR, AT, AU, BD, BE, BG, BH, BO, BR, CA, CH, CL, CN, CO, CY, CZ, DE, DK, DZ, EC, EG, ES, FI, FR, GB, GH, GR, GT, HK, HR, HU, ID, IE, IL, IN, IR, IS, IT, JO, JP, KE, KR, LB, LU, LV, MA, MT, MX, MY, NG, NL, NO, NP, NZ, OM, PA, PE, PH, PK, PL, PT, PY, QA, RO, SA, SE, SG, SI, TH, TR, TW, UA, VE, VN, ZA
Quantity3,752 systems
Serial numbers100058561, 100058562, 100058565, 100058572, 100058573, 100058574, 100058576, 100058835, 100058840, 100058844, 100058845, 100058846, 100058902, 100058903, 100058904, 100058905, 100058906, 100058908, 100058911, 100058914, 100058917, 100058968, 100058970, 100058971, 100058975 and 475 more
Model numbers1076709, V60

Product

Philips Respironics V60 Ventilator Part Number 1076709 - Product Usage: intended to be used only with various combinations of Respironics-recommended patient circuits, interfaces (masks), humidifiers, and other accessories.

Reason for recall

Potential issues with ventilator navigation ring interface and liquid ingress: 1) The navigation ring may become inoperative or operate intermittently; 2) The navigation ring may cause ventilator parameter values to change erratically during setting entries and adjustments.