AMSCO 7053HP Washer/Disinfector Model/Catalog Number: 7053HP Product… — Class II medical device recall Z-0910-2026
Ongoing recall by Steris Corporation, reported 2025-12-17, distributed nationwide. Wire connected to the electrical box may shift out of its intended position, which may result in electrical ar
| Recall number | Z-0910-2026 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Steris Corporation, Mentor, OH, United States |
| Recall initiated | 2025-11-19 |
| Classified | 2025-12-11 |
| Reported by FDA | 2025-12-17 |
| Distributed | Nationwide (US) |
| Countries outside the US | AU |
| Quantity | 22 |
| UPC / GTIN / UDI-DI | 00724995176655, 00724995176662, 00724995176679 |
| Model numbers | 7053HP |
Product
AMSCO 7053HP Washer/Disinfector Model/Catalog Number: 7053HP Product Description: The AMSCO 7053HP Washer/Disinfector is intended for use in the cleaning and intermediate-level disinfection of soiled reusable simple hard-surfaced rigid surgical instruments (such as forceps and clamps), utensils (such as bedpans, urinals, trays, bowl, basin and kidney dishes), rubber and plastic goods, theatre shoes and other similar and related articles found in healthcare facilities.
Reason for recall
Wire connected to the electrical box may shift out of its intended position, which may result in electrical arcing. Electrical arcing remains internal within the unit (no risk of shock to users).
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0910-2026. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.