Data Foundry

Medical device recalls 2015

The Siemens Symbia T series is intended for use by appropriately… — Class II medical device recall Z-0912-2015

Terminated recall by Siemens Medical Solutions USA, Inc., reported 2015-01-14. Siemens Medical Solutions USA, Inc. has received reports of unintended radial motion of the detectors. This un

Recall numberZ-0912-2015
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmSiemens Medical Solutions USA, Inc., Hoffman Estates, IL, United States
Recall initiated2014-08-12
Classified2015-01-06
Reported by FDA2015-01-14
Terminated2016-12-23
Countries outside the USAE, AR, AT, AU, BD, BE, BG, BR, CA, CH, CL, CN, CO, CR, CZ, DE, DK, DZ, EC, EE, EG, ES, FI, FR, GB, GR, HK, HR, HU, ID, IE, IL, IN, IR, IT, JO, JP, KR, KW, LT, LV, MA, MX, MY, NL, NO, NZ, OM, PE, PH, PK, PL, PT, QA, RO, SA, SE, SG, SI, SK, TH, TN, TR, TW, UA, VE, VN, ZA
Quantity1,330 systems
Serial numbers1000, 1001, 1002, 1003, 1004, 1005, 1006, 1007, 1008, 1009, 1010, 1011, 1012, 1013, 1014, 1015, 1016, 1017, 1018, 1019, 1020, 1021, 1022, 1023, 1024 and 379 more

Product

The Siemens Symbia T series is intended for use by appropriately trained health care professionals to aid in detecting, localizing, diagnosing, staging and restaging of lesions, tumors, disease and organ function for the evaluation of diseases and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The images produced by the system can also be used by the physician to aid in radiotherapy treatment planning and interventional radiology procedures. SPECT: The SPECT component is intended to detect or image the distribution of radionuclides in the body or organ (physiology), using the following techniques: planar imaging, whole body imaging, and tomographic imaging for isotopes with energies up to 588 keV. CT: The CT component is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data (anatomy) from either the same axial plane taken at different angles or spiral planes taken at different angles. SPECT+CT: The SPECT and CT components used together acquire SPECT/CT images. The SPECT images can be corrected for attenuation with the CT images, and can be combined (image registration) to merge the patients physiological (SPECT) and anatomical (CT) images. Software: The syngo MI Applications software is a display and analysis package intended to aid the clinician in the assessment and quantification of pathologies in images produced from SPECT, PET, CT, and other imaging modalities.

Reason for recall

Siemens Medical Solutions USA, Inc. has received reports of unintended radial motion of the detectors. This unintended motion, should it occur, may cause the patient compression related injuries.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0912-2015. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.