Data Foundry

Medical device recalls 2025

GE Healthcare Carestation anesthesia system Power Management Board — Class I medical device recall Z-0915-2026

Ongoing recall by GE Medical Systems China Co., Ltd., reported 2025-12-24, distributed nationwide. Potential unexpected shutdown of Carestation 600 and 700 Series Anesthesia Systems containing certain power ma

Recall numberZ-0915-2026
Product typeMedical device
ClassificationClass I
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmGE Medical Systems China Co., Ltd., Wuxi, China
Recall initiated2025-11-14
Classified2025-12-18
Reported by FDA2025-12-24
DistributedNationwide (US)
Countries outside the USAE, AR, AU, BD, BG, BR, BS, CA, CN, CR, CZ, DE, DO, EC, EG, ES, FI, FR, GB, GT, HK, ID, IE, IL, IN, IQ, IT, JP, KR, KW, LK, LT, LV, MA, MX, MY, NG, NL, NO, NZ, PA, PE, PH, PK, PL, PT, QA, RO, RS, RU, SA, SE, SG, SI, SK, SV, TH, TN, TW, UA, VE, VN, ZA
Quantity203 units

Product

GE Healthcare Carestation anesthesia system Power Management Board (PMB) Field Replaceable Unit (FRU), 2076139-001-S

Reason for recall

Potential unexpected shutdown of Carestation 600 and 700 Series Anesthesia Systems containing certain power management boards, if the AC mains power is unplugged or in the event of an AC mains power failure. Anesthesia systems only operate on battery power in a rare event that AC mains power is lost and there is no continuous backup emergency power. If AC power is interrupted, the system will not automatically switch over to the battery supply mode and will reboot. If this issue occurs, a temporary disruption of mechanical ventilation, manual ventilation, and volatile agent delivery may occur. Following the reboot, the system will not return to the previous ventilation settings.

Machine access