Data Foundry

Medical device recalls 2015

Lotus TAVR 27mm — Class I medical device recall Z-0917-2015

Terminated recall by Boston Scientific Corporation, reported 2015-01-21. Lotus valve became unlocked during release from the delivery system. This may lead to percutaneous insertion o

Recall numberZ-0917-2015
Product typeMedical device
ClassificationClass I
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmBoston Scientific Corporation, Los Gatos, CA, United States
Recall initiated2014-11-19
Classified2015-01-13
Reported by FDA2015-01-21
Terminated2015-03-26
Countries outside the USCH, DE, ES, FI, FR, GB, IT, NO, SE
Quantity278 units
Serial numbers14076030, 14076031, 14079012, 14079013, 14149023, 14149031, 14149038, 14149040, 14150028, 14150029, 14153080, 14153081, 14153082, 14154040, 14157024, 14157025, 14157026, 14157027, 14157029, 14157030, 14157031, 14157032, 14157033, 14165041, 14165042 and 16 more
Model numbersLTV27

Product

Lotus TAVR 27mm, Transcatheter Aortic Valve Prosthesis Premounted on Delivery System; Sterile Material number H749LTV270, Catalog Number LTV27; Product Usage: The Lotus Valve System is intended to improve aortic valve function for symptomatic subjects with severe calcific aortic stenosis (aortic valve area [AVA] of <1.0 cm2 or AVA index of <0.6 em /m2) who are at extreme or high risk for standard surgical valve replacement.

Reason for recall

Lotus valve became unlocked during release from the delivery system. This may lead to percutaneous insertion of another valve, resulting in a Valve in Valve (ViV) implantation. If this occurs, the first valve could embolize and it may be necessary to convert the patient to surgery.