Data Foundry

Medical device recalls 2022

MM-TSM — Class III medical device recall Z-0917-2022

Ongoing recall by Volcano Corp, reported 2022-04-20, distributed in AK, AL, AR, AZ, CA, CO, CT, DC…. Multi-Modality Touch Screen Modules (MM-TSM), used with integrated IVUS systems, may crack, or become damaged,

Recall numberZ-0917-2022
Product typeMedical device
ClassificationClass III
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmVolcano Corp, San Diego, CA, United States
Recall initiated2022-02-03
Classified2022-04-12
Reported by FDA2022-04-20
DistributedAK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY
Quantity1,358
Reason classespackaging
Model numbers300002018101, 300002018102, 797403

Product

MM-TSM (Touch Screen Module/Secondary Controller/Bedside Controller), Part Numbers: 300002018101, 300002018102, used with IntraSight Integrated (Part Number: 797403) IVUS systems

Reason for recall

Multi-Modality Touch Screen Modules (MM-TSM), used with integrated IVUS systems, may crack, or become damaged, which may affect their usability. They could become non-functional. Therefore, bottom protective beams will be replaced with full protective metal brackets.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0917-2022. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.