Data Foundry

Medical device recalls 2014

B — Class II medical device recall Z-0921-2014

Terminated recall by B. Braun Medical, Inc., reported 2014-02-12. Internal testing of two lots of the AccuFlo CT-2000-100 Elastomeric Infusion Pump identified flow rates which

Recall numberZ-0921-2014
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmB. Braun Medical, Inc., Allentown, PA, United States
Recall initiated2013-12-13
Classified2014-02-03
Reported by FDA2014-02-12
Terminated2014-09-03
Model numbers3H0125F151, 3H0225F151, CT-2000-100LOTS
Expiration dates2018-07

Product

B. Braun AccuFlo Elastomeric Infusion Pump 200ML/H 100 ML Generally used for delivery of antibiotic and local anesthesia medications.

Reason for recall

Internal testing of two lots of the AccuFlo CT-2000-100 Elastomeric Infusion Pump identified flow rates which are outside the specified infusion time and may result in overly rapid infusion of medications.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0921-2014. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.