ARCHITECT c4000 Cuvette Segment — Class II medical device recall Z-0921-2018
Terminated recall by Abbott Laboratories, Inc, reported 2018-03-14, distributed nationwide. There is a potential to generate falsely depressed patient results in the cuvettes adjacent to the detached se
| Recall number | Z-0921-2018 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Abbott Laboratories, Inc, Irving, TX, United States |
| Recall initiated | 2017-12-08 |
| Classified | 2018-03-06 |
| Reported by FDA | 2018-03-14 |
| Terminated | 2019-07-11 |
| Distributed | Nationwide (US) |
| Countries outside the US | AE, AF, AR, AT, AU, BD, BE, BG, BR, BS, CA, CH, CL, CN, CO, CR, CW, CZ, DE, DK, DO, DZ, EC, EE, EG, ES, FI, FR, GB, GR, GT, HK, HN, HR, HU, ID, IE, IL, IN, IQ, IT, JM, JO, KE, KR, KW, KY, LB, LT, LU, LV, MA, MX, MY, NG, NL, NO, NZ, OM, PA, PE, PH, PK, PL, PT, PY, QA, RO, RS, RU, SA, SE, SG, SI, SK, SV, TH, TN, TR, TT, TW, UY, VE, VN, ZA |
| Quantity | 3,634 architect |
| Reason classes | potency, device malfunction |
| Model numbers | 02P75-01/02P24 |
Product
ARCHITECT c4000 Cuvette Segment, Part #02P75-01/02P24, a component to the ARCHITECT Clinical Chemistry system. There are 9 segments with each cuvette segment holding 11 cuvettes for a total of 99 cuvettes per reaction carousel. The firm name on the label is Abbott Laboratories, Abbott Park, IL. The Abbott ARCHITECT Clinical Chemistry Systems are designed to perform automated chemistry tests utilizing photometry and potentiometric technology. The cuvette segments are a component of the system and are racks that sit in the reaction carousel and hold cuvettes.
Reason for recall
There is a potential to generate falsely depressed patient results in the cuvettes adjacent to the detached segment post for any assay run on the c4000, c8000, and c16000 instrument.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0921-2018. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.