Data Foundry

Medical device recalls 2018

ARCHITECT c4000 Cuvette Segment — Class II medical device recall Z-0921-2018

Terminated recall by Abbott Laboratories, Inc, reported 2018-03-14, distributed nationwide. There is a potential to generate falsely depressed patient results in the cuvettes adjacent to the detached se

Recall numberZ-0921-2018
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmAbbott Laboratories, Inc, Irving, TX, United States
Recall initiated2017-12-08
Classified2018-03-06
Reported by FDA2018-03-14
Terminated2019-07-11
DistributedNationwide (US)
Countries outside the USAE, AF, AR, AT, AU, BD, BE, BG, BR, BS, CA, CH, CL, CN, CO, CR, CW, CZ, DE, DK, DO, DZ, EC, EE, EG, ES, FI, FR, GB, GR, GT, HK, HN, HR, HU, ID, IE, IL, IN, IQ, IT, JM, JO, KE, KR, KW, KY, LB, LT, LU, LV, MA, MX, MY, NG, NL, NO, NZ, OM, PA, PE, PH, PK, PL, PT, PY, QA, RO, RS, RU, SA, SE, SG, SI, SK, SV, TH, TN, TR, TT, TW, UY, VE, VN, ZA
Quantity3,634 architect
Reason classespotency, device malfunction
Model numbers02P75-01/02P24

Product

ARCHITECT c4000 Cuvette Segment, Part #02P75-01/02P24, a component to the ARCHITECT Clinical Chemistry system. There are 9 segments with each cuvette segment holding 11 cuvettes for a total of 99 cuvettes per reaction carousel. The firm name on the label is Abbott Laboratories, Abbott Park, IL. The Abbott ARCHITECT Clinical Chemistry Systems are designed to perform automated chemistry tests utilizing photometry and potentiometric technology. The cuvette segments are a component of the system and are racks that sit in the reaction carousel and hold cuvettes.

Reason for recall

There is a potential to generate falsely depressed patient results in the cuvettes adjacent to the detached segment post for any assay run on the c4000, c8000, and c16000 instrument.