Data Foundry

Medical device recalls 2018

ARCHITECT c8000 Cuvette Segment — Class II medical device recall Z-0922-2018

Terminated recall by Abbott Laboratories, Inc, reported 2018-03-14, distributed nationwide. There is a potential to generate falsely depressed patient results in the cuvettes adjacent to the detached se

Recall numberZ-0922-2018
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmAbbott Laboratories, Inc, Irving, TX, United States
Recall initiated2017-12-08
Classified2018-03-06
Reported by FDA2018-03-14
Terminated2019-07-11
DistributedNationwide (US)
Countries outside the USAE, AF, AR, AT, AU, BD, BE, BG, BR, BS, CA, CH, CL, CN, CO, CR, CW, CZ, DE, DK, DO, DZ, EC, EE, EG, ES, FI, FR, GB, GR, GT, HK, HN, HR, HU, ID, IE, IL, IN, IQ, IT, JM, JO, KE, KR, KW, KY, LB, LT, LU, LV, MA, MX, MY, NG, NL, NO, NZ, OM, PA, PE, PH, PK, PL, PT, PY, QA, RO, RS, RU, SA, SE, SG, SI, SK, SV, TH, TN, TR, TT, TW, UY, VE, VN, ZA
Quantity3,357 architect
Reason classespotency, device malfunction
Model numbers01G46-01/01G06

Product

ARCHITECT c8000 Cuvette Segment, Part #01G46-01/01G06, a component to the ARCHITECT Clinical Chemistry system. There are 11 segments with each cuvette segment holding 15 cuvettes for a total of 165 cuvettes per reaction carousel. The firm name on the label is Abbott Laboratories, Abbott Park, IL. The Abbott ARCHITECT Clinical Chemistry Systems are designed to perform automated chemistry tests utilizing photometry and potentiometric technology. The cuvette segments are a component of the system and are racks that sit in the reaction carousel and hold cuvettes.

Reason for recall

There is a potential to generate falsely depressed patient results in the cuvettes adjacent to the detached segment post for any assay run on the c4000, c8000, and c16000 instrument.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0922-2018. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.