Data Foundry

Medical device recalls 2024

BrightView X — Class I medical device recall Z-0930-2024

Ongoing recall by Philips North America, reported 2024-02-14, distributed nationwide. Detector may unexpectedly fall due to a component failure. If the detector is positioned below center of gantr

Recall numberZ-0930-2024
Product typeMedical device
ClassificationClass I
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmPhilips North America, Cambridge, MA, United States
Recall initiated2023-12-15
Classified2024-02-02
Reported by FDA2024-02-14
DistributedNationwide (US)
Reason classesdevice malfunction
Model numbers6-DIGIT

Product

BrightView X, gamma camera for SPECT; Model Nos.: 6-digit format 882478; 12-digit format 453560824741 453560829261;

Reason for recall

Detector may unexpectedly fall due to a component failure. If the detector is positioned below center of gantry, there is a potential for abrasion, contusion, laceration, and/or fracture to the patient's lower limbs, as well as an interruption to normal system operation. If detector is positioned above center of gantry, there may be an interruption to normal system operation.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0930-2024. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.