Orthofix NorthStar OCT Navigation Surgical Technique Guide — Class II medical device recall Z-0937-2026
Ongoing recall by SEASPINE ORTHOPEDICS CORPORATION, reported 2025-12-24, distributed in AL, AZ, CA, CO, CT, FL, IL, IN…. The NorthStar OCT Navigation Surgical Technique Guide (STG) incorrectly identified both Medtronic s Solera and
| Recall number | Z-0937-2026 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | SEASPINE ORTHOPEDICS CORPORATION, Carlsbad, CA, United States |
| Recall initiated | 2025-11-17 |
| Classified | 2025-12-17 |
| Reported by FDA | 2025-12-24 |
| Distributed | AL, AZ, CA, CO, CT, FL, IL, IN, MI, MO, MS, NJ, NV, NY, OK, OR, PA, RI, SC, SD, TX, UT, VA, WA |
| Countries outside the US | AU, CL, MX |
| Quantity | 310 |
| UPC / GTIN / UDI-DI | 10889981260033, 10889981260040, 10889981260057, 10889981260064, 10889981260071, 10889981260088, 10889981260095, 10889981260101, 10889981260118, 10889981260125, 10889981275495 |
| Lot numbers | AL658549C, AL658549CRRR, AL658641C, AL661470C, AL671917D, AL683456F, AL702785F, BK104899AR, BK104899ARR, BK104899ARRR, BK104899ARRR4, BK104900AR, BK104900ARRR, BK104901ARRR, BK104902ARRR, BK104903ARRR, BK104904ARRR, BK106822A, BK106825A, BK106827A, BK106829A, BK106831A, BK106831ARRR, BK106833A, BK107806A, BK107806ARRR, BK108366A, BK108366ARRR, BK131159C, BK131160C, CR49020D, CR49021C, CR49023D, CR49060D, CR49061C, CR49062C, CR49620C, CR49621C, CR49822D, CR50158D and 49 more |
| Model numbers | PC2-400012, PC2-400013, PC2-400135, PC2-400140, PC2-400155, PC2-400235, PC2-400240, PC2-400245, PC2-400345, PC2-400355 |
Product
Orthofix NorthStar OCT Navigation Surgical Technique Guide, and instruments: Navigation Driver, REF: PC2-400012; Navigation Straight Bone Probe, REF: PC2-400013; Navigation, 3.5mm Drill, REF: PC2-400135; Navigation, 4.0mm Drill, REF: PC2-400140; Navigation, 4.5mm Drill, PC2-400145; Navigation, 5.5mm Drill, REF: PC2-400155; Navigation, 3.5mm Tap, REF: PC2-400235; Navigation, 4.0mm Tap, REF: PC2-400240; Navigation, 4.5mm Tap, REF: PC2-400245; Navigation, 4.5mm Dual Lead Tap, REF: PC2-400345; Navigation, 5.5mm Dual Lead Tap, REF: PC2-400355
Reason for recall
The NorthStar OCT Navigation Surgical Technique Guide (STG) incorrectly identified both Medtronic s Solera and Medtronic s Infinity system toolcards as compatible with the NorthStar OCT Navigation Instruments. However, the NorthStar OCT Navigation Instruments are compatible for use with Medtronic s Solera toolcards only.