Data Foundry

Medical device recalls 2017

TVC Imaging System model TVC-MC9 — Class II medical device recall Z-0940-2017

Terminated recall by Infraredx, Inc., reported 2017-01-11. Near-infrared spectroscopic image (Chemogram) generated by the TVC Imaging System TVC-MC9 in some cases demons

Recall numberZ-0940-2017
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmInfraredx, Inc., Burlington, MA, United States
Recall initiated2016-12-15
Classified2017-01-03
Reported by FDA2017-01-11
Terminated2018-10-09
Countries outside the USKR
Quantity7 units
Serial numbers12718M00009, MC9-15005, MC9-15006, MC9-15007, MC9-15008, MC9-15012, MC9-15013
Model numbersTVC-MC9

Product

TVC Imaging System model TVC-MC9 , for the near-infrared examination of coronary arteries in patients undergoing invasive coronary angiography The TVC Imaging System is an intravascular imaging device that simultaneously assesses vessel composition and structure using near-infrared spectroscopy (NIRS) and intravascular ultrasound (IVUS). This dual-modality instrument performs near-infrared spectroscopic analysis of the vessel to detect lipid core-containing plaques of interest (LCP) displayed in a map called a Chemogram, and simultaneously generates high resolution IVUS images that display structural details of the vessel and plaque in transverse and longitudinal views

Reason for recall

Near-infrared spectroscopic image (Chemogram) generated by the TVC Imaging System TVC-MC9 in some cases demonstrated an unusual appearance in that the yellow in the image (indicating the presence of lipid core plaques) was horizontally-oriented

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0940-2017. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.