TRU-Balance 3 Power Positioning Systems — Class II medical device recall Z-0942-2017
Terminated recall by Pride Mobility Products Corp, reported 2017-01-11, distributed nationwide. The interface between the TB3 back and tilt base can separate when tilted beyond the specified angle.
| Recall number | Z-0942-2017 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Pride Mobility Products Corp, Exeter, PA, United States |
| Recall initiated | 2016-12-01 |
| Classified | 2017-01-03 |
| Reported by FDA | 2017-01-11 |
| Terminated | 2020-03-19 |
| Distributed | Nationwide (US) |
| Quantity | 12,343 chairs |
Product
TRU-Balance 3 Power Positioning Systems, TB3 Lift/Tilt/Recline (Gen 1). Mobility Medical Device
Reason for recall
The interface between the TB3 back and tilt base can separate when tilted beyond the specified angle.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0942-2017. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.