Data Foundry

Medical device recalls 2019

AF w/AS 550 x 1-14 ml/hr — Class II medical device recall Z-0953-2019

Terminated recall by Arrow International Inc, reported 2019-03-06, distributed nationwide. Possible cracks on the body of the fill port connector, which may result in a leak. If the leak is identified

Recall numberZ-0953-2019
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmArrow International Inc, Reading, PA, United States
Recall initiated2018-10-02
Classified2019-02-25
Reported by FDA2019-03-06
Terminated2021-11-12
DistributedNationwide (US)
Quantity2
Reason classespackaging

Product

AF w/AS 550 x 1-14 ml/hr, 5ml30 LOCPNB, Product code MVBX30XLCPNB, Arrow AutoFuser Disposable Pain Control Pump

Reason for recall

Possible cracks on the body of the fill port connector, which may result in a leak. If the leak is identified while the user is filling the pump, there may be a delay while an alternative pump is located. If the leak goes undetected it may result in a reduced level of drug delivery to the patient and consequently a lower level of pain relief than that prescribed or expected. This may result in the need for other, additional analgesic drug administration.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0953-2019. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.