Data Foundry

Medical device recalls 2021

BioGX SARS-CoV-2 Reagents for BD MAX System Catalog number — Class II medical device recall Z-0954-2021

Terminated recall by Becton Dickinson & Co., reported 2021-02-17, distributed nationwide. Some N1 and N2 master mix pouches packaged within the lots of BioGX SARS-CoV-2 Reagents for BD MAX System may

Recall numberZ-0954-2021
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmBecton Dickinson & Co., Sparks, MD, United States
Recall initiated2021-01-07
Classified2021-02-05
Reported by FDA2021-02-17
Terminated2021-12-03
DistributedNationwide (US)
Quantity938 kits
Reason classeslabeling, packaging
UPC / GTIN / UDI-DI00382904442132
Lot numbersK20-507, K20-508, K20-509, K20-516, K20-517, K21-020, K21-021
Model numbers444213

Product

BioGX SARS-CoV-2 Reagents for BD MAX System Catalog number (REF) 444213. In vitro diagnostic. SARS-CoV-2 Reagents is a real-time RT-PCR test intended for the qualitative detection of nucleic acid from the SARS-CoV-2 in nasopharyngeal, nasal, mid-turbinate, and oropharyngeal swab specimens, nasopharyngeal wash/aspirate or nasal aspirates obtained from individuals suspected of COVID-19 by their healthcare provider.

Reason for recall

Some N1 and N2 master mix pouches packaged within the lots of BioGX SARS-CoV-2 Reagents for BD MAX System may be mislabeled. This could result in a kit containing two pouches of N1 master mix or two pouches of N2 master mix, instead of one pouch of each. If laboratory personnel do not notice that both pouches contain tubes with the same color foil seal and use two of the same tubes, there is a low potential for a false negative SARS-CoV-2 result. A false negative SARS-CoV-2 result could lead to an increased risk of disease progression for the patient and additional spread of SARS-CoV-2 if the patient is not quarantined.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0954-2021. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.