Medtronic MiniMed 640G Insulin Infusion Pump — Class I medical device recall Z-0958-2020
Ongoing recall by Medtronic Inc., reported 2020-02-19, distributed in AK, AL, AR, AZ, CA, CO, CT, DC…. There have been reported incidents of a loose reservoir that can no longer be locked into the pump. The reserv
| Recall number | Z-0958-2020 |
|---|---|
| Product type | Medical device |
| Classification | Class I |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Medtronic Inc., Northridge, CA, United States |
| Recall initiated | 2019-11-21 |
| Classified | 2020-02-07 |
| Reported by FDA | 2020-02-19 |
| Distributed | AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY |
| Countries outside the US | AT, AU, BE, BR, CA, CH, CO, CZ, DE, DK, ES, FI, FR, GB, GR, HU, IE, IL, IT, JP, KR, MX, NL, NO, NZ, PL, SA, SE, SK |
| Model numbers | MMT-1511, MMT-1512, MMT-1711, MMT-1712, MMT-1751, MMT-1752 |
Product
Medtronic MiniMed 640G Insulin Infusion Pump, Ref#s/Model(s) - MMT-1511, MMT-1711, MMT-1512, MMT-1712, MMT-1751, and MMT-1752 *Not for US Distribution
Reason for recall
There have been reported incidents of a loose reservoir that can no longer be locked into the pump. The reservoir can become loose due to a broken or missing retainer ring that prevents a proper lock. If the reservoir is not properly locked into the pump, it could lead to over or under delivery of insulin, which could then result in hypoglycemia or hyperglycemia.