Data Foundry

Medical device recalls 2020

Medtronic MiniMed 640G Insulin Infusion Pump — Class I medical device recall Z-0958-2020

Ongoing recall by Medtronic Inc., reported 2020-02-19, distributed in AK, AL, AR, AZ, CA, CO, CT, DC…. There have been reported incidents of a loose reservoir that can no longer be locked into the pump. The reserv

Recall numberZ-0958-2020
Product typeMedical device
ClassificationClass I
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmMedtronic Inc., Northridge, CA, United States
Recall initiated2019-11-21
Classified2020-02-07
Reported by FDA2020-02-19
DistributedAK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY
Countries outside the USAT, AU, BE, BR, CA, CH, CO, CZ, DE, DK, ES, FI, FR, GB, GR, HU, IE, IL, IT, JP, KR, MX, NL, NO, NZ, PL, SA, SE, SK
Model numbersMMT-1511, MMT-1512, MMT-1711, MMT-1712, MMT-1751, MMT-1752

Product

Medtronic MiniMed 640G Insulin Infusion Pump, Ref#s/Model(s) - MMT-1511, MMT-1711, MMT-1512, MMT-1712, MMT-1751, and MMT-1752 *Not for US Distribution

Reason for recall

There have been reported incidents of a loose reservoir that can no longer be locked into the pump. The reservoir can become loose due to a broken or missing retainer ring that prevents a proper lock. If the reservoir is not properly locked into the pump, it could lead to over or under delivery of insulin, which could then result in hypoglycemia or hyperglycemia.