PRYTIME MEDICAL REBOA Catheter Introducer Kit — Class II medical device recall Z-0959-2021
Terminated recall by Prytime Medical Devices, Inc., reported 2021-02-17, distributed nationwide. There is a potential that a catheter convenience set contains a 21G access needle instead of the correct 18G n
| Recall number | Z-0959-2021 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Prytime Medical Devices, Inc., Boerne, TX, United States |
| Recall initiated | 2020-12-14 |
| Classified | 2021-02-05 |
| Reported by FDA | 2021-02-17 |
| Terminated | 2022-09-28 |
| Distributed | Nationwide (US) |
| Lot numbers | 3590 |
| Model numbers | KT1835C |
Product
PRYTIME MEDICAL REBOA Catheter Introducer Kit, REF: KT1835C, Contents: Needle, 18 GA x 7 CM ECHOGENIC, STERILE, 1 EA, SYRINGE, 30CC, STERILE 1 EA, INTRODUCER SHEATH KIT, 7 FR W/.035 SHORT GUIDEWIRE, STERILE 1 EA., PREFILLED, STERILE FIELD SALINE SYRINGE, 10 ML, 4 EA, CLAMP, 5 FR, STERILE, 1 EA, SCALPEL, #11SS, STERILE, 1 EA, SUTURE, NYLON, 18", 2-0, REVERSE CUTTING, STERILE 1 EA, THREE QUARTER DRAPE, STERILE 1 EA, INSERT, SALINE 1 EA, INSERT, CLAMP 1 EA UDI: (01) 0 0863092 00010 1
Reason for recall
There is a potential that a catheter convenience set contains a 21G access needle instead of the correct 18G needle. If the 21G needle is used to gain vascular access then the guidewire included in the kit will not pass through the access needle to facilitate insertion of an introducer sheath and could therefore cause procedure delay or injury.