Alinity ci-series System Control Module — Class II medical device recall Z-0961-2026
Ongoing recall by Abbott Laboratories, reported 2025-12-24, distributed nationwide. Abbott Laboratories is recalling their Alinity ci-series System Control Module, a chemistry/immunoassay anayze
| Recall number | Z-0961-2026 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Abbott Laboratories, Irving, TX, United States |
| Recall initiated | 2025-11-13 |
| Classified | 2025-12-18 |
| Reported by FDA | 2025-12-24 |
| Distributed | Nationwide (US) |
| Countries outside the US | AE, AR, AT, AU, BD, BE, BG, BH, BR, BS, CA, CH, CL, CN, CO, CR, CY, CZ, DE, DK, DZ, EC, EE, EG, ES, FI, FR, GB, GR, GT, HK, HN, HR, HU, ID, IE, IL, IN, IQ, IR, IT, JM, JO, JP, KE, KR, KW, KY, LB, LT, LV, MA, MT, MX, MY, NI, NL, NO, NP, NZ, OM, PA, PE, PH, PK, PL, PT, PY, QA, RO, RS, RU, SA, SE, SG, SI, SK, SV, TH, TN, TR, TT, TW, UA, UY, VN, ZA |
| Quantity | 11,653 |
| Reason classes | software |
| UPC / GTIN / UDI-DI | 00380740137410, 06952999490048 |
| Serial numbers | SCM01010, SCM01012, SCM01016, SCM01035, SCM01043, SCM01051, SCM01063, SCM01069, SCM01075, SCM01076, SCM01078, SCM01082, SCM01085, SCM01086, SCM01087, SCM01088, SCM01092, SCM01093, SCM01094, SCM01095, SCM01098, SCM01099, SCM01101, SCM01102, SCM01106 and 475 more |
| Model numbers | 03R70-01, 03R70-20 |
Product
Alinity ci-series System Control Module (SCM), REF: 03R70-01, 03R70-20
Reason for recall
Abbott Laboratories is recalling their Alinity ci-series System Control Module, a chemistry/immunoassay anayzer, by correction. The reason for the recall is potential performance issues found in the Alinity ci-series System software versions 3.6.1. and lower that could lead to erroneous results for multiple analytes. The issue was identified by Abbott during the internal testing of complaint investigations.