da Vinci 5 — Class II medical device recall Z-0964-2026
Ongoing recall by Intuitive Surgical, Inc., reported 2025-12-31, distributed nationwide. Surgical system Error 23062 due to connector1) Partially damaged "Potential problem with ergonomics: Disable -
| Recall number | Z-0964-2026 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Intuitive Surgical, Inc., Sunnyvale, CA, United States |
| Recall initiated | 2024-08-21 |
| Classified | 2025-12-19 |
| Reported by FDA | 2025-12-31 |
| Distributed | Nationwide (US) |
| Quantity | 144 |
| UPC / GTIN / UDI-DI | 00886874119747 |
| Serial numbers | 10476314, 10485693, 10496049, 10497992, 10503432, 10503433, 10507302, 10509551, 10511044, 10511047, 10516467, 10518960, 10518961, 10522438, 10522439, 10522441, 10523771, 10523773, 10523775, 10523776, 10523777, 10523778, 10523779, 10523780, 10523781 and 119 more |
| Model numbers | 380730-40 |
Product
da Vinci 5, Surgeon Console Column Motor Connector; ASSY,DV5 CONSOLE,SSC,IS5000; Part Number: 380730-40
Reason for recall
Surgical system Error 23062 due to connector1) Partially damaged "Potential problem with ergonomics: Disable - Retry " or 2) Failed "Disable Ergonomics: Viewer, headrest, armrest, and pedal tray position cannot be adjusted until da Vinci is restarted. Contact Technical Support to report the issue: Cancel-Disable" ergonomics locked until motor replaced; may lead to delay/aborted/alternate procedure