Spacelabs BleaseSirius Anesthesia Workstation — Class I medical device recall Z-0968-2013

Terminated recall by Del Mar Reynolds Medical, Ltd., reported 2013-03-27, distributed nationwide. A defect in CAS I/II Absorbers in the Spacelabs Anesthesia Workstations and Service Kits may cause a condition

Recall numberZ-0968-2013
Product typeMedical device
ClassificationClass I
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmDel Mar Reynolds Medical, Ltd., Hertford, N/A, United Kingdom
Recall initiated2013-01-15
Classified2013-03-20
Reported by FDA2013-03-27
Terminated2013-05-14
DistributedNationwide (US)
Countries outside the USAR, CN, MX, TW
Serial numbersABSO-004448, ABSO-004458, ABSO-004459, ABSO-004460, ABSO-004461, ABSO-004462, ABSO-004463, ABSO-004465, ABSO-004466, ABSO-004467, ABSO-004468, ABSO-004469, ABSO-004470, ABSO-004471, ABSO-004472, ABSO-004473, ABSO-004474, ABSO-004475, ABSO-004476, ABSO-004477, ABSO-004478, ABSO-004479, ABSO-004480, ABSO-004481, ABSO-004482 and 83 more
Model numbers050-0659-00, 050-0901-00, 12200902

Product

Spacelabs BleaseSirius Anesthesia Workstation, BleaseFocus Anesthesia Workstation, and CAS I/II Absorber contained in Service Kits Part Number 12200902, 050-0659-00 and 050-0901-00 and any workstation installed with the CAS I/II Absorber. Anesthesia workstation for use in the hospital environment and locations not requiring portability.

Reason for recall

A defect in CAS I/II Absorbers in the Spacelabs Anesthesia Workstations and Service Kits may cause a condition leading to an increase in the carbon dioxide concentration within the inspired gas being delivered to the patient.