Data Foundry

Medical device recalls 2016

EndoVive 3s Low Profile Balloon Kits Part Number: M00548750 — Class II medical device recall Z-0977-2016

Terminated recall by Xeridiem Mediem Medical Devices Inc, reported 2016-03-02, distributed nationwide. A feeding (reflux) valve leakage issue. The feeding valve appears to close inconsistently, or close in a delay

Recall numberZ-0977-2016
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmXeridiem Mediem Medical Devices Inc, Tucson, AZ, United States
Recall initiated2015-12-23
Classified2016-02-25
Reported by FDA2016-03-02
Terminated2016-10-25
DistributedNationwide (US)
Quantity39
Reason classespackaging
Lot numbers929716
Model numbersM00548750

Product

EndoVive 3s Low Profile Balloon Kits Part Number: M00548750 (XMD P/N 70-0050-A18)

Reason for recall

A feeding (reflux) valve leakage issue. The feeding valve appears to close inconsistently, or close in a delayed manner, when an extension set is disconnected from the Low Profile Balloon device, thus allowing feeding solution or gastric contents to leak out of the device in some instances.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0977-2016. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.