Data Foundry

Medical device recalls 2015

ADVIA Chemistry XPT interfaced to the CentraLink Data Management… — Class II medical device recall Z-0987-2015

Terminated recall by Siemens Healthcare Diagnostics, Inc., reported 2015-01-21. Enabling sending of preliminary/initial results on the ADVIA Chemistry XPT creates a risk that a critical resu

Recall numberZ-0987-2015
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmSiemens Healthcare Diagnostics, Inc., Newark, DE, United States
Recall initiated2014-12-16
Classified2015-01-15
Reported by FDA2015-01-21
Terminated2016-10-27
Quantity7
Model numbers10817209, 10818262

Product

ADVIA Chemistry XPT interfaced to the CentraLink Data Management System V14x in specific configurations: The CentraLink system software is a network solution provider and multi-system data manager for the instruments and lab automation systems (LAS) within the lab. The CentraLink software consolidates data from all connected instruments so that an operator can review and edit patient and quality control results from a single location.

Reason for recall

Enabling sending of preliminary/initial results on the ADVIA Chemistry XPT creates a risk that a critical result could be overwritten by the same result rather than showing the repeated result on CentraLink. A result could be erroneous and critical and appear to be verified upon repeat.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0987-2015. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.