ADVIA Chemistry XPT interfaced to the CentraLink Data Management… — Class II medical device recall Z-0987-2015
Terminated recall by Siemens Healthcare Diagnostics, Inc., reported 2015-01-21. Enabling sending of preliminary/initial results on the ADVIA Chemistry XPT creates a risk that a critical resu
| Recall number | Z-0987-2015 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Siemens Healthcare Diagnostics, Inc., Newark, DE, United States |
| Recall initiated | 2014-12-16 |
| Classified | 2015-01-15 |
| Reported by FDA | 2015-01-21 |
| Terminated | 2016-10-27 |
| Quantity | 7 |
| Model numbers | 10817209, 10818262 |
Product
ADVIA Chemistry XPT interfaced to the CentraLink Data Management System V14x in specific configurations: The CentraLink system software is a network solution provider and multi-system data manager for the instruments and lab automation systems (LAS) within the lab. The CentraLink software consolidates data from all connected instruments so that an operator can review and edit patient and quality control results from a single location.
Reason for recall
Enabling sending of preliminary/initial results on the ADVIA Chemistry XPT creates a risk that a critical result could be overwritten by the same result rather than showing the repeated result on CentraLink. A result could be erroneous and critical and appear to be verified upon repeat.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0987-2015. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.