Data Foundry

Medical device recalls 2023

Cardinal Health Jackson-Pratt Channel Drain — Class II medical device recall Z-0989-2023

Ongoing recall by Cardinal Health 200, LLC, reported 2023-01-25, distributed nationwide. Cardinal Health recently received multiple complaints regarding various performance issues with Cardinal Healt

Recall numberZ-0989-2023
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmCardinal Health 200, LLC, Waukegan, IL, United States
Recall initiated2022-12-14
Classified2023-01-18
Reported by FDA2023-01-25
DistributedNationwide (US)
Countries outside the USCA, HK, SA, SG
Quantity8,563 units
UPC / GTIN / UDI-DI20885380182638, 50885380182639
Lot numbersN210278, N210347, N210368, N210423, N210460, N210468, N210469, N210543, N220050, N220099, N220139, N220163
Model numbersJP-2231

Product

Cardinal Health Jackson-Pratt Channel Drain, 19 Fr, REF JP-2231

Reason for recall

Cardinal Health recently received multiple complaints regarding various performance issues with Cardinal Health Jackson-Pratt Channel Drains including lack of radiopacity in wound drains, dull trocars impacting product performance, lack of flexibility of wound drains, and broken wound tips.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0989-2023. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.