Cardinal Health Jackson-Pratt Channel Drain — Class II medical device recall Z-0991-2023
Ongoing recall by Cardinal Health 200, LLC, reported 2023-01-25, distributed nationwide. Cardinal Health recently received multiple complaints regarding various performance issues with Cardinal Healt
| Recall number | Z-0991-2023 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Cardinal Health 200, LLC, Waukegan, IL, United States |
| Recall initiated | 2022-12-14 |
| Classified | 2023-01-18 |
| Reported by FDA | 2023-01-25 |
| Distributed | Nationwide (US) |
| Countries outside the US | CA, HK, SA, SG |
| Quantity | 1,920 units |
| UPC / GTIN / UDI-DI | 20885380182652, 50885380182653 |
| Lot numbers | N210278, N210368, N210423, N210460, N210468, N210543, N220050, N220099, N220139, N220163 |
| Model numbers | JP-2233 |
Product
Cardinal Health Jackson-Pratt Channel Drain, 15 Fr, REF JP-2233
Reason for recall
Cardinal Health recently received multiple complaints regarding various performance issues with Cardinal Health Jackson-Pratt Channel Drains including lack of radiopacity in wound drains, dull trocars impacting product performance, lack of flexibility of wound drains, and broken wound tips.