Data Foundry

Medical device recalls 2014

Superstand stand-up wheelchair HPS-2 A standup wheelchair is a device… — Class II medical device recall Z-0993-2014

Terminated recall by The Standing Company, reported 2014-02-26, distributed nationwide. Power Superstand Standing Wheel chair, flex shaft (the flexible wand which supports the joystick control modul

Recall numberZ-0993-2014
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmThe Standing Company, Saginaw, MI, United States
Recall initiated2013-05-16
Classified2014-02-19
Reported by FDA2014-02-26
Terminated2014-06-25
DistributedNationwide (US)
Quantity119
Reason classesdevice malfunction
Serial numbers1018, 1306, 1313, 1342, 1349, 1358, 1359, 1362, 1391, 1392, 1399, 1402, 1406, 1407, 1426, 1427, 1441, 1451, 1453, 1477, 1480, 1482, 1494, 1497, 1505 and 93 more

Product

Superstand stand-up wheelchair HPS-2 A standup wheelchair is a device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position. The device incorporates an external manually controlled mechanical system that is intended to raise a paraplegic to an upright position by means of an elevating seat.

Reason for recall

Power Superstand Standing Wheel chair, flex shaft (the flexible wand which supports the joystick control module) may fracture if it is bent repeatedly and /or beyond 90 degrees. A newly designed Flexshaft which corrects this problem is available at no charge.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0993-2014. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.